肺定位放射線治療におけるカーボンファイバカウチの線量影響
富永 弘史荒木 不次男下東 吉信兼武 渚冨山 友希川﨑 啓一岩下 雄紀坂田 潤一奥田 智子
著者情報
ジャーナル フリー

2013 年 69 巻 4 号 p. 400-406

詳細
抄録

The aim of this study was to measure the dose attenuation caused by a carbon fiber radiation therapy table (Imaging Couch Top; ICT, BrainLab) and to evaluate the dosimetric impact of ICT during stereotactic body radiation therapy (SBRT) in lung tumors. The dose attenuation of ICT was measured using an ionization chamber and modeled by means of a treatment planning system (TPS). SBRT was planned with and without ICT in a lung tumor phantom and ten cases of clinical lung tumors. The results were analyzed from isocenter doses and a dose-volume histogram (DVH): D95, Dmean, V20, V5, homogeneity index (HI), and conformity index (CI). The dose attenuation of the ICT modeled with TPS agreed to within ±1% of the actually measured values. The isocenter doses, D95 and Dmean with and without ICT showed differences of 4.1–5% for posterior single field and three fields in the phantom study, and differences of 0.6–2.4% for five fields and rotation in the phantom study and six fields in ten clinical cases. The dose impact of ICT was not significant for five or more fields in SBRT. It is thus possible to reduce the dose effect of ICT by modifying the beam angle and beam weight in the treatment plan.