頭部CT撮影における頭部固定具が画質と線量に及ぼす影響
横町 和志西丸 英治藤岡 知加子木口 雅夫石風呂 実船間 芳憲粟井 和夫
著者情報
ジャーナル フリー

2014 年 70 巻 10 号 p. 1166-1172

詳細
抄録

Purpose: For emergency or pediatric head CT scans, a simplified pillow made of hard sponge instead of a dedicated head holder may be used if it is difficult to immobilize the head. However, the radiation dose when using a simplified head holder may be increased due to radiation absorption by the patient couch if the automatic exposure control (AEC) system is used. In this phantom study, we compared the radiation dose delivered when using a dedicated and a simplified head holder. Materials and Methods: We used a dedicated-type and a pillow-type head holder made of hard sponge (simplified head holder). We placed a 20 cm-diameter cylindrical phantom made of water-equivalent material and an anthropomorphic head phantom in the head holders and then scanned them five times with a 64-detector CT scanner (VCT, GE Healthcare). We performed step-and-shoot and helical scanning with AEC; the noise index was set to 2.8. We measured the radiation dose using fluorescent glass dosimeters in the head phantom and the image noise at five sites in the cylindrical phantom. All values were averaged. Results: With step-and-shoot scans, the mean image noise with the dedicated and the simplified head holder was 3.30 ± 0.05 [SD] and 3.20 ± 0.05, respectively. With helical scans they were 3.00 ± 0.09 and 2.88 ± 0.03, respectively. There was no statistically significant difference (p = 0.02 and 0.04, Student’s t-test). The radiation doses with the dedicated and the simplified head holder were 58.6 and 70.4 mGy, respectively, for step-and-shoot scanning and 41.8 and 49.0 mGy, respectively, for helical scanning. The doses were thus significantly higher with the simplified head holder for both step-and-shoot and helical scanning (p < 0.01 and < 0.01). Conclusion: We recommend the use of a dedicated head holder for head scanning with AEC since the radiation dose was lower than with the simplified head holder.

著者関連情報
© 2014 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top