薬剤供給システムの異なる18F-FDG PET/CT 検査における臨床画像の比較
我妻 慧三輪 建太宮司 典明村田 泰輔梅田 拓朗大澤 敦滝口 智洋小泉 満
著者情報
ジャーナル フリー

2014 年 70 巻 4 号 p. 339-345

詳細
抄録

Objective: PET images are affected by scanner model, reconstruction conditions, injected dose, scan duration, patient health status and FDG radiopharmaceutical supply systems. The present study compares images of 40 patients using commercial and in-house FDG systems with one PET scanner (Aquiduo). Methods: The PET images were evaluated using the physical indexes of NECpatient, NECdensity and SNRliver proposed by the Japanese guidelines for oncological FDG-PET/CT, and by visual assessment. Results: There were no significant differences in the physical indexes between PET images generated using commercial and in-house FDG. The physical indexes were also acceptable according to the recommended Japanese guidelines. NECdensity was higher when a higher dose/body weight of commercial FDG was injected (correlation coefficient: r=0.576, p<0.001) and lower when BMI was lower and in-house FDG was injected (r=−0.786, p<0.0001). These results suggest that scan duration should be increased if the injected dose of commercial FDG/body weight is <5.5 MBq/kg, and if individuals with BMI >21.4 kg/m2 are injected with in-house FDG. Conclusions: Scan duration should be varied depending on FDG supply systems to ensure more accurate image quality and quantitative values during evaluations of response to therapy and prognostic prediction.