日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
臨床技術
多施設における循環器撮影装置のIVR基準点での線量率の比較
石橋 徹今田 直幸山下 由香利麻生 弘哉松本 頼明稲田 智沖野 瑞穂野中 春輝水谷 宏
著者情報
ジャーナル フリー

2012 年 68 巻 3 号 p. 260-268

詳細
抄録

The management of the radiation dose is very important in interventional radiology (IVR), especially in percutaneous coronary intervention (PCI). Therefore, we measured entrance surface doses at the interventional reference point of 27 cardiac intervention procedures in 22 cardiac catheterization laboratories around Hiroshima, and compared these doses. Recently, for cardiac interventional radiology, the X-ray machines using flat-panel detectors (FPD) instead of image intensifiers (I.I.) is increasing; 13 systems used FPD and 14 systems used I.I. For fluoroscopy rate, the difference between laboratories was 9 times. For cineangiography rate, the difference between laboratories was 7 times. In addition, between both devices, the I.I. group is bigger than the FPD group. When comparing by the same condition, for the dose at the interventional reference point, no significant difference was detected between the FPD group and the I.I. group. This study shows that FPD is not available for reducing the radiation dose simply. Therefore, it is necessary that we think of the balance with image quality and radiation dose. The optimization of the devices and cardiac intervention procedures becomes very important.

著者関連情報
© 2012 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top