直接変換型FPDを搭載したディジタルX線透視撮影システムの画質特性と視覚評価の対応
坂口 太郎片山 礼司杜下 淳次坂井 真二黒木 英郁大久保 征司前田 孝早渕 尚文
著者情報
ジャーナル フリー

2010 年 66 巻 11 号 p. 1457-1466

詳細
抄録

The purpose of this study was to examine the correlation between the basic imaging properties of two digital radiographic X-ray systems with a direct conversion flat-panel detector and their image qualities, which were evaluated by the observer in hard copy and soft copy studies. The subjective image quality was evaluated and compared in terms of the low-contrast detectability and image sharpness in the two digital radiographic X-ray systems. We applied the radiographs of a contrast detail phantom to the evaluation of low-contrast detectability and analyzed the contrast detail diagrams. Finally, low-contrast detectability was evaluated by the image quality figure (IQF) calculated from the contrast detail diagrams. Also, the subjective image sharpness of human dry bones of two systems was examined and evaluated by the normalized-rank method. The results indicated that System A tended to provide superior subjective image quality compared to System B in both observer studies. We also found high correlations between IQFs and basic imaging properties, such as the noise power spectrum (NPS) and the noise equivalent quantum (NEQ). In conclusion, the low-contrast detectability of the two digital radiographic X-ray systems with a direct conversion flat-panel detector corresponded to the NPS and the NEQ in both outputs (soft copy and hard copy). On the other hand, the subjective image sharpness of human dry bones was affected by their noise properties.

著者関連情報
© 2010 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top