The Influence of Posture Change on Measurements of Relative Body Fat in the Bioimpedance Analysis Method
Shinichi DemuraShunsuke YamajiFumio GoshiYoshinori Nagasawa
Author information
JOURNAL FREE ACCESS

2001 Volume 20 Issue 1 Pages 29-35

Details
Abstract

The purpose of this study was to clarify the influence of posture change on relative body fat in the bioelectrical impedance analysis (BIA) method. The subjects were 30 Japanese healthy young adult males (age: 19.8 ± 1.4 years, height: 172.3 ± 5.8 cm, weight: 67.1 ± 8.2 kg). We used devices with different body segment inductions, between the hand and foot (H-F BIA) and between hands (H-H BIA), and set four measurement conditions differing in posture (supine or sitting), during rest and measurement. The reliabilities of %BF in the H-H and H-F BIA methods were very high (r=0.995, 0.966), and the relationship in %BF between the UW method and each BIA method was mid-range (r=0.767, 0.709). Although there were no differences in %BF among different measurement postures in the H-F BIA method, %BF in the H-H BIA method increased significantly when the posture was changed just before measurement. This indicated that it is necessary to pay attention to the posture change just before measurement in the H-H BIA method.